Model 187 & 189
True RMS Multimeter
Users Manual
August 2000, Rev.2, 6/02
© 2000, 2002 Fluke Corporation. All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies.
Zobrazit stránku 0
1 2 3 4 5 6 ... 95 96

Shrnutí obsahu

Strany 1 - Model 187 & 189

®Model 187 & 189True RMS MultimeterUsers ManualAugust 2000, Rev.2, 6/02© 2000, 2002 Fluke Corporation. All rights reserved. Printed in U.S.A.All p

Strany 2 - Lifetime Limited Warranty

Model 187 & 189Users Manualviii3-10. Temperature Measurement...

Strany 3 - Table of Contents

1-1Chapter 1Before You StartSafety InformationThe Fluke Model 187 and Model 189 True RMSMultimeters (hereafter referred to as the “meter”) complywith:

Strany 4

Model 187 & 189Users Manual1-2Table 1-1. Safety InformationWWarningTo avoid possible electric shock or personal injury, follow these guidelines:•

Strany 5 - Contents (continued)

Before You StartSafety Information11-3Table 1-1. Safety Information (cont.)WWarning• When using the probes, keep your fingers behind the finger guard

Strany 6

Model 187 & 189Users Manual1-4SymbolsSymbols used on the meter and in this manual areexplained in Table 1-2.Table 1-2. International Electrical Sy

Strany 7 - List of Tables

2-1Chapter 2Getting AcquaintedIntroductionAlthough this manual describes the operation of bothModels 187 and 189, all illustrations and examplesassume

Strany 8

Model 187 & 189Users Manual2-2tc031f.epsFigure 2-1. AC Volts DisplayBattery ConsiderationsThe meter uses four AA alkaline batteries. The following

Strany 9 - List of Figures

Getting AcquaintedBattery Considerations22-3Automatic Backlight OffPress T to select the backlight level (low, high, or off.) Inlow or high, the backl

Strany 10

Model 187 & 189Users Manual2-4Rotary SwitchTurn the meter on by selecting any measurement function(identified with white letters around the rotary

Strany 11 - Before You Start

Getting AcquaintedPushbuttons22-5PushbuttonsThe buttons activate features that augment the functionselected with the rotary switch. The buttons are sh

Strany 12 - WWarning

Lifetime Limited WarrantyEach Fluke 20, 70, 80, 170 and 180 Series DMM will be free from defects in material and workmanship for its lifetime. As use

Strany 13

Model 187 & 189Users Manual2-6Table 2-1. Rotary Switch SelectionsPosition Rotary Switch Function % Blue Key Function\AC voltage measurement from

Strany 14

Getting AcquaintedPushbuttons22-7Table 2-1. Rotary Switch Positions (cont.)Position Rotary Switch Function % Blue Key FunctionIAC current measurements

Strany 15 - Getting Acquainted

Model 187 & 189Users Manual2-8Table 2-2. PushbuttonsButton DescriptionYellow ButtonFunction DescriptionNotePress O to access “Yellow Button Functi

Strany 16 - Battery Considerations

Getting AcquaintedPushbuttons22-9Table 2-2. Pushbuttons (cont.)Button DescriptionYellow ButtonFunction Description [• In Setup, increment a digit .•

Strany 17

Model 187 & 189Users Manual2-10Selecting the RangePress R to select either a fixed range or theautorange feature.NoteYou cannot use R in conductan

Strany 18 - Rotary Switch

Getting AcquaintedUnderstanding the Display22-11• dB (in ac volts functions): the dBm or dBV value.• REL: the difference between the present reading

Strany 19 - Pushbuttons

Model 187 & 189Users Manual2-121416 151789101171356243112tc011f.epsFigure 2-4. Display Features

Strany 20

Getting AcquaintedUnderstanding the Display22-13Table 2-3. Display FeaturesNumber Feature Description ~ Continuity test function is selected.

Strany 21

Model 187 & 189Users Manual2-14Table 2-3. Display Features (cont.)Number Feature DescriptionFASTMINMAXAVGFAST MN MX mode is enabled. ( O M)Minim

Strany 22 - AutoHOLD

Getting AcquaintedUnderstanding the Display22-15Table 2-3. Display Features (cont.)Number Feature Description AC+DC For dc volts and dc amps function

Strany 23 - CANCEL

iTable of ContentsChapter Title Page1 Before You Start ...

Strany 24 - Understanding the Display

Model 187 & 189Users Manual2-16Table 2-3. Display Features (cont.)Number Feature Description51000AUTOMANUALRange. Digits display range in use.fM

Strany 25 - Bar Graph

Getting AcquaintedUsing the Input Terminals22-17Using the Input TerminalsAll functions except current use the *andCOM inputs. Current functions use th

Strany 26 - Figure 2-4. Display Features

Model 187 & 189Users Manual2-18Using Display HoldPress Q to enter the Display Hold mode and freezethe present reading and its time stamp. New read

Strany 27 -

Getting AcquaintedUsing AutoHOLD22-19Using AutoHOLDWWarningAutoHOLD mode does not capture unstableor noisy readings. Do not use AutoHOLDmode to determ

Strany 28 - 8.8.8.8.8

Model 187 & 189Users Manual2-20To exit MIN MAX mode, press O N (CANCEL) orturn the rotary switch to a different position. Also, MINMAX mode turns

Strany 29

Getting AcquaintedUsing FAST MN MX22-21Using FAST MN MXFAST MN MX can capture transient signal events as shortas 250 µs, but with decreased accuracy;

Strany 30 - HM88:88

Model 187 & 189Users Manual2-22Using HOLD with MIN MAX or FAST MN MXYou can enable the HOLD mode when in the MIN MAXmode is by pressing Q.No furth

Strany 31 - Using the Input Terminals

3-1Chapter 3Making MeasurementsIntroductionChapter 3 explains how to make measurements. Mostmeasurement functions can be selected by using therotary s

Strany 32

Model 187 & 189Users Manual3-2When measuring voltage, the meter acts like a 10 MΩ(10,000,000 Ω) impedance in parallel with the circuit. Thisloadin

Strany 33 - Using MIN MAX

Making MeasurementsMeasuring Voltage33-3dB Measurements in AC Volts FunctionsThe two ac volts functions allow you to display readings asdeviations in

Strany 34

Model 187 & 189Users ManualiiSelecting the Range...

Strany 35 - Using FAST MN MX

Model 187 & 189Users Manual3-4dB is calculated with the following formula:=VrVxdB10log*20• For dBm, Vr is the voltage across the referencer

Strany 36 - Using Relative Mode (REL)

Making MeasurementsMeasuring Voltage33-5AC over DC DC over AC AC + DCtc024f.epsFigure 3-3. AC and DC DisplayWhen the meter shows ac over dc or dc over

Strany 37 - Making Measurements

Model 187 & 189Users Manual3-6MIN MAXHOLDREL%msHzRANGEdBdBac+dcac+dcac+dcac+dcFnSmAmAAmVVmVVOFFCAAAAmACOMVTEMPERATUREAMIN MAXHOLDREL%msHzRANGEdBdB

Strany 38 - Measuring AC Voltage

Making MeasurementsMeasuring Resistance33-7MIN MAXHOLDREL%msHzRANGEdBdBac+dcac+dcac+dcac+dcFnSmAmAAmVVmVVOFFCAAAAmACOMVTEMPERATUREAMIN MAXHOLDREL%msHz

Strany 39

Model 187 & 189Users Manual3-8Keep the following in mind when measuring resistance:• Because the meter’s test current flows through allpossible p

Strany 40 - Measuring DC Voltage

Making MeasurementsUsing Conductance for High Resistance Tests33-9Continuity testing provides you with both a visualindication of the state encountere

Strany 41 - AC over DC DC over AC AC + DC

Model 187 & 189Users Manual3-10MIN MAXHOLDREL%msHzRANGEdBdBac+dcac+dcac+dcac+dcFnSmAmAAmVVmVVOFFCAAAAmACOMVTEMPERATUREAMIN MAXHOLDREL%msHzRANGEdBd

Strany 42 - Measuring Resistance

Making MeasurementsUsing Conductance for High Resistance Tests33-11MIN MAXHOLDREL%msHzRANGEdBdBac+dcac+dcac+dcac+dcFnSmAmAAmVVmVVOFFCAAAAmACOMVTEMPERA

Strany 43 - LOGGING MULTIMETER

Model 187 & 189Users Manual3-12Measuring CapacitanceCautionTo avoid possible damage to the meter or tothe equipment under test, disconnect circuit

Strany 44 - Testing for Continuity

Making MeasurementsTesting Diodes33-13MIN MAXHOLDREL%msHzRANGEdBdBac+dcac+dcac+dcac+dcFnSmAmAAmVVmVVOFFCAAAAmACOMVTEMPERATUREAMIN MAXHOLDREL%msHzRANGE

Strany 45

Contents (continued)iiiMeasuring AC Current ... 3-18Measuri

Strany 46 - Figure 3-6. Continuity Test

Model 187 & 189Users Manual3-14MIN MAXHOLDREL%msHzRANGEdBdBac+dcac+dcac+dcac+dcFnSmAmAAmVVmVVOFFCAAAAmACOMVTEMPERATUREAMIN MAXHOLDREL%msHzRANGEdBd

Strany 47

Making MeasurementsMeasuring Temperature33-15Measuring TemperatureTo measure temperature, set up the meter as shown inFigure 3-10. The meter begins te

Strany 48 - Measuring Capacitance

Model 187 & 189Users Manual3-16Measuring CurrentWWarningNever attempt an in-circuit currentmeasurement where the open-circuit potentialto earth is

Strany 49 - Testing Diodes

Making MeasurementsMeasuring Current33-173. If you are using the A terminal, set the rotary switch tomA/A. If you are using the mA/µA terminal, set t

Strany 50 - Figure 3-9. Diode Test

Model 187 & 189Users Manual3-18The following are some tips for measuring current:• If the display shows LEAd5 and you are sure themeter is set up

Strany 51 - Measuring Temperature

Making MeasurementsMeasuring Current33-19MIN MAXHOLDREL%msHzRANGEdBdBac+dcac+dcac+dcac+dcFnSmAmAAmVVmVVOFFCAAAAmACOMVTEMPERATUREAMIN MAXHOLDREL%msHzRA

Strany 52 - Measuring Current

Model 187 & 189Users Manual3-20Measuring DC CurrentTo measure dc current, set up the meter as shown inFigure 3-12.You can view separate dc and ac

Strany 53 - Width Modulation

Making MeasurementsMeasuring Current33-21MIN MAXHOLDREL%msHzRANGEdBdBac+dcac+dcac+dcac+dcFnSmAmAAmVVmVVOFFCAAAAmACOMVTEMPERATUREAMIN MAXHOLDREL%msHzRA

Strany 54 - Measuring AC Current

Model 187 & 189Users Manual3-22Measuring FrequencyFrequency is the number of cycles a signal completeseach second. The meter measures the frequenc

Strany 55

Making MeasurementsMeasuring Duty Cycle33-23tc026f.epsFigure 3-14. Hz DisplayThe meter beeps to indicate when a particular pushbuttonis not allowed wh

Strany 56 - Measuring DC Current

Model 187 & 189Users Manualiv6 Maintenance ...

Strany 57

Model 187 & 189Users Manual3-24-Slope Trigger Point+Slope Trigger Point30% Above TriggerPoint100%-Slope Trigger Point+Slope Trigger Point70% Below

Strany 58 - Measuring Frequency

Making MeasurementsMeasuring Pulse Width33-25To measure duty cycle, set up the meter to measurefrequency; then press N a second time. You canselect th

Strany 59 - Measuring Duty Cycle

Model 187 & 189Users Manual3-26+Slope Trigger PointPulseWidthPeriod =Period-Slope Trigger PointPulse WidthStart and Stop on Positive SlopeStart an

Strany 60

Making MeasurementsMeasuring Pulse Width33-27The meter measures pulse width in the 500.00 or1000.0 ms ranges.To measure pulse width, set up the meter

Strany 61 - Measuring Pulse Width

Model 187 & 189Users Manual3-28

Strany 62

4-1Chapter 4Using Memory & Communications FeaturesIntroductionChapter 4 shows you how to use memory andcommunication features available on the met

Strany 63

Model 187 & 189Users Manual4-2Storing Saved ReadingsTo add the current displayed reading to the savedreadings memory, press O R (SAVE).• 5AVEd a

Strany 64

Using Memory & Communications FeaturesViewing Memory Data44-3Viewing Memory DataUse the following procedure to view memory data:NoteViewing memory

Strany 65 - Chapter 4

Model 187 & 189Users Manual4-42734156tc035f.epsFigure 4-1. View DisplayTable 4-1. View DisplayNo. Item Description q arrowicons.Denotes use of [

Strany 66 - Stopping Logging

Using Memory & Communications FeaturesClearing Memory44-5Clearing MemoryYou can clear memory in two ways.• First, If the rotary switch is in the

Strany 67 - Viewing Memory Data

vList of TablesTable Title Page1-1. Safety Information...

Strany 68

Model 187 & 189Users Manual4-6

Strany 69 - Clearing Memory

5-1Chapter 5Changing the Default SettingsIntroductionThe meter allows you to change the default operatingconfiguration of the meter by changing setup

Strany 70

Model 187 & 189Users Manual5-2Table 5-1. Function Specific Setup SelectionsSelection Precondition Option Choices ( ) Factory Default000.0 °Co

Strany 71 - Changing the Default Settings

Changing the Default SettingsSelecting Setup Options55-3Table 5-2. Common Setup SelectionsSelection Option Choices Factory DefaultbEEP Beeper yE5 or n

Strany 72 - to select digit

Model 187 & 189Users Manual5-4Select and edit setup options as follows:• Turn the rotary switch to a measurement function:• Press O T to advance

Strany 73

Changing the Default SettingsAdjusting the Temperature Offset55-5If necessary, adjust the temperature offset until thetemperature on the primary displ

Strany 74

Model 187 & 189Users Manual5-6Selecting Display Resolution(3 1/2 or 4 1/2 Digits)For most functions, you can choose whether the meterdisplays the

Strany 75

Changing the Default SettingsSetting the 24-Hour Clock55-7Setting the 24-Hour ClockThe meter uses 24-hour clock readings as time stampsduring HOLD, Au

Strany 76 - Setting the Power Off Timeout

Model 187 & 189Users Manual5-8Returning to Factory DefaultsYour meter comes with the setup options preset at thefactory. These factory settings ar

Strany 77 - Setting the 24-Hour Clock

6-1Chapter 6MaintenanceIntroductionThis chapter describes basic operator maintenance. Forcalibration and performance test information, order the187 &a

Strany 78 - Saving Setup Options

Model 187 & 189Users Manualvi

Strany 79 - Maintenance

Model 187 & 189Users Manual6-21000V400mAFUSED10A MAXFUSEDCATMIN MAXHOLDREL%msHzRANGEdBdBac+dcac+dcac+dcac+dcFnSmAmAAmVVmVVOFFCAAAAmACOMVTEMPERATUR

Strany 80

MaintenanceReplacing the Batteries66-3Replacing the BatteriesReplace the batteries with four AA batteries (NEDA I5A orIEC LR6).WWarningTo avoid false

Strany 81 - Replacing the Batteries

Model 187 & 189Users Manual6-4F2F1tc037f.epsFigure 6-2. Battery and Fuse Replacement

Strany 82

MaintenanceReplacing the Fuses66-5Replacing the FusesW WarningTo avoid electrical shock or damage to themeter, only use replacement fuses specifiedin

Strany 83 - In Case of Difficulty

Model 187 & 189Users Manual6-6Table 6-1. User-Replaceable PartsDescription Reference Designators Part Number QtyAccess Door, Battery / Fuse MP14 6

Strany 84

7-1Chapter 7SpecificationsSafety and CompliancesMaximum voltage between any terminal and earthground.1000 V dc or rms acCompliances - DUAL RATINGS Com

Strany 85 - Specifications

Model 187 & 189Users Manual7-2Physical SpecificationsDisplay (LCD) Digital: 50000/5000 counts primary display,5000 counts secondary display;update

Strany 86 - Physical Specifications

SpecificationsFeature Summary77-3Feature SummaryFeature DescriptionDual Digital DisplaysAnalog Bar GraphPrimary: 50,000 countsSecondary: 5,000 count

Strany 87 - Feature Summary

Model 187 & 189Users Manual7-4Basic SpecificationsFunction Ranges/DescriptionDC Voltage 0 to 1000 VAC Voltage, true RMS 2.5 mV to 1000 V – 100 kHz

Strany 88 - Basic Specifications

SpecificationsDetailed Accuracy Specifications77-5Detailed Accuracy SpecificationsAccuracy is specified for a period of one year after calibration, at

Strany 89

viiList of FiguresFigure Title Page2-1. AC Volts Display ...

Strany 90

Model 187 & 189Users Manual7-6AccuracyFunction Range Resolution 45-1 kHz 20-45 Hz 1-20 kHz 20 kHz-100 kHzAC µA 500.00 µA 0.01 µA 0.75 % + 20 1 % +

Strany 91

SpecificationsDetailed Accuracy Specifications77-7Accuracy Accuracy Dual Display AC or AC+DC 3Function Range Resolution DC 20 - 45 Hz 45 Hz - 1 kHz 1

Strany 92

Model 187 & 189Users Manual7-8Function Range Resolution AccuracyResistance 1500.00 Ω 0.01 Ω 0.05 % + 10 35.0000 kΩ 0.0001 kΩ 0.05 % + 250.000 kΩ 0

Strany 93

SpecificationsDetailed Accuracy Specifications77-9Function Ranges Resolution AccuracyCapacitance 21.000 nF 0.001 nF 2% + 510.00 nF 0.01 nF 1 % + 5100.

Strany 94

Model 187 & 189Users Manual7-10Function Range Resolution AccuracyFrequency 500.00 Hz 0.01 Hz 1± (0.0050 % + 1)5.0000 kHz 0.0001 kHz50.000 kHz 0.00

Strany 95 - Burden Voltage (A, mA, µA)

SpecificationsFrequency Counter Sensitivity77-11Frequency Counter SensitivityApproximate VAC Sensitivity(RMS Sine Wave)1Input Range15 Hz to100 kHz 250

Strany 96 - Input Characteristics

Model 187 & 189Users Manual7-12Input CharacteristicsFunction Input Impedance (Nominal)Volts, mV 10 MΩ, < 100 pFCommon Mode Rejection Ratio Norm