Manual Supplement Impulse 6000D/7000DP Users
6 4/08
Sine waves ......................................................... 0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz
Pulse................................................................... 30 and 60 BPM, 60 ms pulse width
R-Wave Detection
Waveform ........................................................... Haver-triangle
Amplitude............................................................ 0.05 to 0.45 (by 0.05) mV
0.5 to 5.0 (by 0.5) mV
Rate .................................................................... 30, 60, 80, 120, 200, and 250 BPM
Widths................................................................. 8, 10, 12 ms, and 20 to 200 (by 10) ms
Accuracy............................................................. ±(1 % setting + 1 ms)
Noise Immunity
Wave................................................................... Sine
Line Frequency ................................................... 50 or 60 Hz (± 0.5 Hz)
Amplitude
Range ............................................................. 0.0 to 10.0 (by 0.5) mV
Accuracy......................................................... ±5 %
Transvenous Pacer Pulse Simulation
Widths
Range ............................................................. 0.1, 0.2, 0.5, 1.0, and 2.0 ms
Accuracy......................................................... ±5 % of setting
Amplitude
Range ............................................................. 0 (off) and ±2, ±4, ±6, ±8, ±10, ±12, ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, and
±700 mV