4 - 66 COMMAND REFERENCE
INSTrument:NSELect
INSTrument[:SELect]
Syntax:
INSTrument:NSELect <NRf> | MINimum | MAXimum
INSTrument[:SELect] DIGital | ANALog
<NRf> 1 | 2
1 | MINimum The digital mode (ANALOG key) is activated.
2 | MAXimum The analog mode is activated.
DIGital The digital mode (ANALOG key) is activated.
ANALog The analog mode is activated.
Query form:
INSTrument:NSELect? [MINimum | MAXimum}
Response:
1 | 2
1 The digital mode (ANALOG key) is active. This is
also returned when MAXimum is specified.
2 The analog mode is active. This is also returned
when MINimum is specified.
Query form:
INSTrument[:SELect]?
Response:
DIG | ANAL
DIG The digital mode (ANALOG key) is active.
ANAL The analog mode is active.
Description:
Selects the analog or digital mode of the CombiScope by specifying a predefined
number or name. When one mode is selected, the other mode is deactivated.
After a
*
RST command, the digital mode (ANALOG key) is selected.
Example:
Send
INSTrument:NSELect 2 Analog mode is selected.
Send
INSTrument DIGital Digital mode is selected.
Front panel compliance:
These commands are the remote equivalent of the front panel ANALOG key.